María Marcos-Jubilar2, Inés Martínez-Alfonzo1, Laura Martos3, Carlos Aguilar4, Noelia Vilalta5, Dolors Llobet5, Ana Marco6, David Aragoneses6, Luciana Ricca7, David Ferreira8, Sara Lopes8, João Mariano-Pego9, Ángel Bernardo10, Antonio Moscardó11, Santiago Bonanad11, Susana Gassiot12, Isabel Gutiérrez-Jomarrón13, Jose Antonio Páramo2, Pilar Llamas-Sillero1, Joan Carles Reverter14
1 Hospital Universitario Fundación Jiménez Díaz (Madrid, Spain); 2 Clínica Universitaria de Navarra (Pamplona, Spain); 3 Diagnostica Stago S.A.S. (Asnières-sur-Seine, France); 4 Hospital Santa Bárbara (Soria, Spain); 5 Hospital de la Santa Creu i Sant Pau (Barcelona, Spain); 6 Hospital Universitario General de Alicante (Alicante, Spain); 7 Centro Hospitalar Universitário de São João (Porto, Portugal); 8 Centro Hospitalar Vila Nova de Gaia/Espinho (Vila Nova de Gaia, Portugal); 9 Centro Hospitalar Universitário de Coimbra (Coimbra, Portugal); 10 Hospital Universitario Central de Asturias (Oviedo, Spain); 11 Hospital Universitario La Fe (Valencia, Spain); 12 Hospital Sant Joan de Deu (Barcelona, Spain); 13 Hospital Universitario Príncipe de Asturias (Alcalá de Henares, Spain); 14 Hospital Clínic de Barcelona (Barcelona, Spain).
Corresponding author: Diego Velasco-Rodríguez.
Postal address: Department of Hematology. Hospital Universitario Fundación Jiménez Díaz. Avenida Reyes Católicos 2, 28040 Madrid, Spain. Email: diegovelascorodriguez@gmail.com
ABSTRACT
BACKGROUND
The development of fully-automated thrombin generation (TG) assays represents an opportunity for their potential transition from research to routine.
OBJECTIVES
To conduct an interlaboratory comparability study of fully-automated TG, and to establish accurate normal reference intervals for TG parameters in Spanish and Portuguese population.
METHODS
This observational, prospective and multicentric study was conducted at 9 Spanish and 3 Portuguese centers, in two phases. TG was assessed in platelet-poor plasma (PPP) of healthy subjects with the ST-Genesia® analyzer, using the STG‐ThromboScreen® (STG-TS) reagent. In phase I, three different pools were obtained and 3 aliquots of each were sent to all participants. To study repeatability, an aliquot of each pool was analyzed in triplicate on different days and, for precision, the mean of the triplicates between days was analyzed for each laboratory. Accuracy and bias were determined for the three pools separately. In phase II, TG was assessed in at least 30 individuals for each center.
RESULTS
Excellent repeatability was observed for TG parameters, with the exception of velocity index, and adequate intraday precision for each laboratory. Satisfactory performance (z-score ≤ 2) was observed for 96.2% of TG parameters for groups 1 and 2, and 98% for group 3. Most centers showed acceptable relative error for most parameters. Finally, reference ranges for STG-TS stratified by age and sex were determined.
CONCLUSIONS
TG assessment can be performed in a decentralized manner. The Iberian range will serve as a reference to detect if any aspect of the work protocol needs to be revised.